Single Test Dip Card Xylazine Drug Test Procedure
To use a single panel Xylazine drug test card your will need a
disposable collection cup and follow this procedure:
- Collect urine from the donor in a disposable collection cup.
- If the collection cup has a temperature strip read it and make sure is between temperature range.
- Remove the bottom cap from the single fentanyl drug test card and submerge the drug test card in the urine.
- AVOID passing the arrow when submerging the single Xylazine drug test card since this can flood the test and invalidate the drug test.
- Once all the lines in the C (control) area appear you can read the bottom lines in T (test) area.
- To interpret results in the single Xylazine drug test card follow the image below
Benefits of Our Xylazine Drug Test Card
The Medical Disposables Corp. Xylazine drug test card is an alternative to the traditional laboratory urine drug test. This Xylazine drug test card is based on a specific immunochemical reaction between antibodies and antigens, designed to detect two of its metabolites in the urine. Xylazine is a veterinary analgesic used for pain and well known street names are tranq” and the “zombie drug”.
Xylazine Drug Test Card with Accurate Results
Xylazine is a non-opioid tranquilizer used as a sedative, analgesic, and muscle relaxant in animals such as horses and cattle. In recent years, xylazine has emerged as an adulterant in recreational drugs such as heroin. Combining xylazine with other drugs that cause central nervous system depression compounds the sedative effects and can increase the risk of overdose and death.
Xylazine is a non-opioid added to illicit drugs in an effort to lengthen the primary drugs euphoric effects. Because xylazine is a sedative and not an opioid, it cannot be detected by a standard opioid drug testing panel.
Medical Disposables Corp. is known for their high-quality for urine drug test products, and our new Xylazine drug test card is no exception. It is one of the most sensitive onsite drug tests on the market, designed to detect even the slightest presence Xylazine metabolites.
Xylazine drug test strips detect xylazine in human urine at 50 ng/mL cutoff levels. Test results are ready in 5 minutes and each test cassette shows easy-to-read test results.
Single Urine Test Dipcard Instructions for Use
Step 1
Collect urine from the donor in a disposable collection cup*.
If the collection cup has a temperature strip read it and make sure is between temperature range.
Step 2
Remove the test device from the foil pouch. Remove the cap from the test device. Label the device with patient or control identifications.
Step 3
Inmerse the absorbent end into the urine sample for about 10 - 15 seconds. Make sure that the urine level is not above the marker line printed on the front of the device.
Step 4
Lay the device flat on a clean, dry, non-absorbent surface.
Step 5
Read the result at 5 minutes.
Do not read after 5 minutes.
Interpreting Results
NEGATIVE
|
POSITIVE
|
INVALID
|
|
|
|
Two lines appear. One red line should be in the control region (C), and another apparent red or pink line adjacent should be in the test region (T). This negative result indicates that the drug concentration is below the detectable level.
|
One red line appears in the control region (C). No line appears in the test region (T). This positive result indicates that the drug concentration is above the detectable level.
|
Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test using a new test panel. If the problem persist, discontinue using the lot immediately and contact your manufacturer.
|
WARNING:
THIS TEST DOES NOT EVALUATE DRUG SAFETY OR PURITY.
* Disposable Collection Cup Sold Separately